Chapter 12

Peripheral Devices
N5102A Digital Signal Interface Module

6.

Press the

More (1 of 2)

softkey.

From this softkey menu, select the bit order, swap I and Q, select the polarity of the transmitted
data, and access menus that provide data negation, scaling, gain, offset, and IQ rotation
adjustments.

7.

Press the

Data Negation

softkey.

Negation differs from changing the I and Q polarity. Applied to a sample, negation changes the
affected sample by expressing it in the two's complement form, multiplying it by negative one, and
converting the sample back to the selected numeric format. This can be done for I samples, Q
samples, or both.

The choice to use negation is dependent on the device being tested and how it needs to receive
the data.

8.

Press the

Gain, Offset & Scaling

softkey.

Use the softkeys in this menu for the following functions:

reduce sample values for both I and Q using the

Scaling

softkey

increase or decrease the sample values independently for I and Q using the

I Gain

and

Q Gain

softkeys

compensate for or add a DC offset using the

I Offset

and

Q Offset

softkeys

rotate the data on the IQ plane using the

Rotation

softkey

9.

Make any required scaling, gain, offset, or rotation adjustments to properly test the device.

10. Press

Return

>

Return

to return to the first- level softkey menu.

Configuring the Clock Signal

1.

Refer to

Figure 12- 13

. Press the

Clock Setup

softkey.

Figure 12-13

Clock Setup Menu Location

Accesses the Clock Setup Menu