Chapter 4

129

4

Optimizing Performance

In the following sections, this chapter describes procedures that improve the performance of the
Agilent PSG signal generator.

“Using the ALC” on page 129

“Using External Leveling” on page 130

“Creating and Applying User Flatness Correction” on page 133

“Using the Option 521 Detector Calibration (Option 521)” on page 146

“Adjusting Reference Oscillator Bandwidth (Option UNR/UNX/UNY)” on page 147

“Optimizing Phase Noise and Harmonics Below 3.2 GHz (Option UNX/UNY)” on page 147

Using the ALC

Selecting ALC Bandwidth

CAUTION

Internal leveling works from –15 dBm to maximum power.

For internal leveling, the signal generator uses automatic leveling control (ALC) circuitry prior to the
RF output. ALC bandwidth has five selections: automatic, 100 Hz, 1 kHz, 10 kHz, and 100 kHz. In
automatic mode (the preset selection), the signal generator automatically selects the ALC bandwidth
depending on the configuration and settings (see

Figure 4- 1

).

Figure 4-1

Decision Tree for Automatic ALC Bandwidth Selection

RF OUTPUT

< 2 MHz

AM OFF

PULSE OFF

AM OFF

PULSE ON

AM ON

PULSE ON

AM ON

PULSE OFF

ALC BW

100 Hz

ALC BW

1 kHz

ALC BW

10 kHz

No

No

No

Yes

Yes

Yes

Yes

ALC BW

100

kHz

ARB On

Yes

No

No

Yes