Chapter 3

109

Basic Digital Operation

Using Waveform Markers

Using the RF Blanking Marker Function

While you can set a marker function (described as

Marker Routing

on the softkey label) either before or

after setting the marker points (

page 104

), setting a marker function before you set marker points

may change the RF output. RF Blanking includes ALC hold (described on

page 98

, note

Caution

regarding unleveled power). The signal generator blanks the RF output when the marker signal goes
low.

1.

Using the factory–supplied segment SINE_TEST_WFM, set Marker 1 across points 1

180 (

page 104

).

2.

From the

Marker Utilities

menu (

page 101

), assign RF Blanking to Marker 1:

Press

Marker Routing

>

Pulse/RF Blank

>

Marker 1

.

RF Output

Marker pulse on the Event 1 signal.