Delete Lot
1. Open the Protocols page, then open the Protocols tab. Click the Stds & Ctrls tab.
2. In the Installed Kits And Lots section, click the lot you want to delete, then click Delete.
Export Lot
NOTE: Lots and kits can only be exported if the protocol they were originally
created with exists within the system (i.e. if the protocol has been
deleted, the lot or kit cannot be exported).
1. Open the Protocols page, then open the Protocols tab. Click the Stds & Ctrls tab.
2. In the Installed Kits And Lots section, click the lot you want to export, then click Export.
The Save As dialog box opens.
3. Navigate to the location you wish to export the file to, then click Save.
Edit Lot
1. Open the Protocols page, then open the Protocols tab. Click the Stds & Ctrls tab.
2. In the Installed Kits And Lots section, select a lot and then click Edit. The Std/Ctrl
Details tab opens. Change the lot information as appropriate.
Performing System Maintenance
Performing Individual Maintenance Commands
Open the Maintenance page, then click the Auto Maint tab to display the automated
maintenance options. To perform an individual maintenance command, click on the
appropriate button in the Cmds & Routines tab. For a description of each of these
commands, see
WARNING:
Lack of proper maintenance can lead to inaccurate test results
and failure of system components. Please follow all scheduled
maintenance routines to ensure proper functioning of
equipment.
Perform System Calibration
Perform system calibration weekly. In addition, recalibrate the system if the delta cal
temperature exceeds +/-3ºC. For further information, see
.
Manage Maintenance Routines
You manage maintenance routines at the Cmds & Routines tab of the Maintenance page.
Open the Maintenance page, then click the Cmds & Routines tab.
Create New Routine
Create maintenance routines to facilitate your startup procedure, shutdown procedure,
troubleshooting, or calibration. Ensure that the maintenance routine you create is consistent
For In Vitro Diagnostic Use
Using the Software
89