Revive after storage routine – Luminex xPONENT for MAGPIX 4.2 User Manual
Page 24

NOTE: When you adjust and save the probe height settings for all three
areas under a plate name, all areas retain the adjustment.
WARNING:
Correct sample probe height is critical to successful sample
acquisition and calibration. Problems with the sample probe can
lead to fluid leaks and inhibit sample acquisition.
CAUTION:
Ensure that the probe height is set correctly before calibrating
the system.
FIGURE 1.
Sample Probe Height Adjustment
Revive After Storage Routine
NOTE: The Revive After Storage routine is necessary when the system
runs for the first time and is recommended when the system has
been idle for more than a week.
After you have adjusted the sample probe height, run the Revive After Storage (Luminex)
routine.
1. Open the Maintenance page, then the Cmds & Routines tab.
2. Select Revive After Storage (Luminex) from the drop-down list. The Revive After
Storage routine performs the following commands:
• Prime
• Rinse
• Alcohol Flush
• Rinse