Luminex 100 User Manual Version 1.7 User Manual
Page 85
x
MAP Technology
Maintenance and Cleaning
PN 89-00002-00-063 Rev. A
7 - 5
2. On the back of the Luminex 100 analyzer, remove the thumb-
screw at the top of the air intake filter access door.
3. Pull the access door off of the Luminex 100 analyzer.
4. Pull the filter out from the unit, about three or four inches.
5. Grasp the tubing. Remove the filter with one hand, and hold the
tubing with the other hand.
6. Connect a new filter to the tubing.
7. Put the filter inside the panel, then reattach the panel door.
Figure 7-2. Rear of Luminex 100 Analyzer, Air Filter Exposed
To replace the Luminex XYP Air Intake Filter:
1. Disconnect the Luminex XYP from ac power by turning off the
power switch on the rear of the instrument, then unplug the
power cord from the wall source.
2. On the back of the instrument, to the left side, gently remove the
screen from the filter. Do not remove the screws.
Figure 7-3. Back of Luminex XYP, Removing Air Filter