Fluidic – Luminex 100 User Manual Version 1.7 User Manual
Page 30

Luminex 100 User Manual Version 1.7
x
MAP Technology
3 - 6
PN 89-00002-00-063 Rev. A
Fluidic
Sample Arm: Transports the sample from the sample plate to the
cuvette. Upon operation, the carriage drops automatically to the
sample plate for sample retrieval.
Sample Probe: Acquires sample from the sample plate.
Cheminert
®
fitting: Holds the sample probe in place. Use the
wheel to adjust the sample probe height.
Alignment Guide: Directs the sample probe into the Luminex
XYP instrument.
Figure 3-2. Sample Uptake Components
Access doors: There are two access doors on the face of the
analyzer. The center door provides access to the syringe. The left
door provides service panel access. Inside the left door, a bracket
holds the heater block for the Luminex XYP. There is also a clip for
the allen wrench included with the Luminex 100 analyzer.
Warning:
During operation,
this system contains exposed,
moving parts which can cause
puncture wounds. Risk of per-
sonal injury is present. Keep
hands and fingers away from the
sample probe.
Note:
The probe shield is not in
place for this photo. You should
always replace the probe shield
before operating the system.
Alignment
Guide
Cheminert
Fitting
Vertical Adjustment
Screw