Luminex xyp specifications, Recommended equipment – Luminex 100 User Manual Version 1.7 User Manual
Page 28
Luminex 100 User Manual Version 1.7
x
MAP Technology
3 - 4
PN 89-00002-00-063 Rev. A
Luminex XYP
Specifications
General
•
Operating temperature: 15°C to 30°C (59°F to 86°F)
•
Humidity: 20% to 80%, noncondensing
•
Altitude: designed to operate at 2400 m (7874 ft.) above mean
sea level or below.
•
Physical dimensions: 44 cm (17.25 in) W x 61 cm (24 in) D x 8
cm (3 in) H
•
Weight: 14.5 kg (32 lbs)
•
UL installation category: UL Installation Category II, as defined
in Annex J of UL 3101-1.
•
Pollution degree: UL Pollution Degree 2, as defined in section
3.7.3.2 of UL 3101-1.
•
Heater operating range: 35°C to 60°C (95°F to 140°F) with
tolerance 0°C to +2°C.
•
Shipping and storage: allowable shipping and storage
temperature and humidity ranges, 0°C to +50°C and 20%-80%
noncondensing, respectively.
Power
•
Input voltage range: 100-240 V~, 1.8 Amps, 47-63 Hz
•
AC Inlet Fuse: 3 Amp, 250 V~, Fast Acting
Electronics
Communications interface: RS-232
Recommended
Equipment
You may find the following items helpful to protect the electronic
components of your Luminex 100 System.
Printer
We recommend that you have a printer installed to your system so
that you can print reports and diagnostics. We recommend an HP
LaserJet 2200 printer.
Uninterruptible power
supply (UPS)
Luminex highly recommends using an uninterruptible power supply
(UPS) to protect your system from a power outage. Choose a supply
that can provide 1050 Watts for at least 45 minutes. The UPS should
be UL listed and CSA certified. The UPS also needs to be CE
marked if used internationally.
Surge Protector
If you do not use a UPS, use a surge protector. Choose a protector
that meets your needs. Factors to consider include electrical
environment, endurance, suppressed voltage rating, and method of
protection. It should have six outlets, be rated at least 1500 Watts,
and be UL listed, CSA certified, CE marked for nondomestic use
when used internationally.