MAP Technology
System Startup
PN 89-00002-00-063 Rev. A
5 - 3
To add a new lot of xMAP classification calibration
microspheres:
1. Click Calibrate. The Start Calibration dialog box opens.
Figure 5-2. Start Calibration Dialog Box
2. Click New in the Classification Channel box. The Add New
Classification Calibrator dialog box opens.
Figure 5-3. Add New Classification Calibrator Dialog Box
3. Enter the information into the dialog box, using the values on the
Certificate of Quality that came with your xMAP calibration
microspheres. You do not need to enter a value for CL3;
however, you must click within the CL3 box and press Enter to
save the values.
4. Click OK. You can now use this classification channel xMAP
calibration microsphere lot to calibrate your instrument.