Run a single sample, Interrupt a run – Luminex 100 User Manual Version 1.7 User Manual
Page 77

x
MAP Technology
Data Acquisition
PN 89-00002-00-063 Rev. A
6 - 27
9. Remove the microtiter plate from the plate holder.
10. Click Retract from the toolbar to retract the plate holder.
Run a Single Sample
To run a single sample:
1. Click Eject from the toolbar to eject the plate holder.
2. Place a 96-well microtiter plate on the plate holder.
3. Click Retract from the toolbar to retract the plate holder.
4. On the results table part of the acquisition screen, select the sam-
ple that you want to read.
5. Check the Single checkbox below the start button on the main
control screen.
6. Click Start on the main control screen. The Luminex 100 reads
just the selected sample.
7. Click Eject from the toolbar to eject the plate holder.
8. Remove the microtiter plate from the plate holder.
9. Click Retract from the toolbar to retract the plate holder.
Interrupt a Run
During Sample
Acquisition
To interrupt a run while the system acquires a sample:
1. Click Cancel. Sampling stops after you click Cancel, and a win-
dow pops up asking if you want to acquire the next sample. If
you click Yes, sampling proceeds with the next sample. Click No
to stop the session completely.
2. To continue acquiring sample, highlight the sample with which
you want to resume sampling with and click Start.
Between Samples
The Single checkbox becomes accessible for several seconds
between samples. Click the Single checkbox. The software then
pauses between acquiring the next sample run.
Click OK to continue acquiring the next sample.