Laurel Electronics LAUREATE SERIES 2 COUNTER-TIMER-SERIAL INPUT METER User Manual

Page 55

background image

55

Additional programmable features of the serial input meter are detailed in the “Command
Mode for Remote Display Operation of Counter / Timer” and “Data Formats” sections of the
Custom ASCII Protocol Serial Communications Manual. In particular, Mode 12 (hex C), which
is invoked by setting the first digit under ConFIG to C, allows extraction of data from an ASCII
string that contains multiple data values or non-numeric characters. This mode can accom-
modate selected Start and Stop characters. Any number of characters between the Start
character and the data can be masked OFF. Up to 8 display characters (including sign and
decimal point) can be masked ON. Any number of characters between the last displayed
character and the Stop character can be masked OFF.

Instrument Setup software is required to set up parameters for the Remote Display in Mode 12
(hex C). This software is downloadable from our website.

SELECTED FRONT PANEL SETUP ITEMS FOR SERIAL INPUT METER (not consecutive)

If the MENU

key does not work, see Section 9 “Enabling & Locking Out Menu Items.”

Press
Menu

Press Digit
Select Key

Press Value Select
Key

SEtuP

Setup

_000_0

Control inputs 1 and 2

F

1 = Tare enable, 2 = Tare

Control input 1 must be at 0V or grounded for
Tare to operate.

ConFiG

Configu-
ration

__0000

Display mode

6

Remote display (H, K, L commands)

7

Single-value remote display

8

Show 1

st

string value, slaved to another meter

9

Show 2

nd

string value, slaved to another meter

A

Show 3

rd

string value, slaved to another meter

B

Show 4

th

string value, slaved to another meter

C

Custom Start, Stop, Skip, Show characters

.

ti_Out

.

Time-out

_000.00 _000.00 _000.00
_000.00 _000.00

Select digit to flash.

Select

0

thru

9

for flashing digit to set time-out in

seconds. Decimal point location is fixed for 10 ms
resolution.

.

SEr 1

.

__000U

Baud rate

Fixed parameters:
No parity, 8 data bits,
1 stop bit

0

300 baud

1

600 baud

2

1200 baud

3

2400 baud

4

4800 baud

5

9600 baud

6

19200 baud

.

SEr 2

.

__0000U

Meter address

Select

0

thru

F

for addresses 1 thru 15. Select

0.

thru

F.

(with decimal point) for addresses 16 thru 31.

.

SEr 3

.

_00000U

RS485

0

Full duplex

1

Half duplex

___000U

Serial protocol

0

Custom ASCII

1

Modbus RTU

2

Modbus ASCII

SEr 4

.

___000U

Parity

0

None

1

Odd

2

Even

_

Addr

___000 __000 ___000

Modbus address

__158

Select

0

through

9

for flashing digit. Address

range is 1 to 247.