beautypg.com

Laurel Electronics LAUREATE SERIES 2 COUNTER-TIMER-SERIAL INPUT METER User Manual

Page 35

background image

35

Press
Menu

Press Digit
Select Key

Press Value Select
Key

ConFiG

Configura-
tion

__0000

Display mode

0

Normal, overload to exponential format

1

Normal, overload to 999999

2

1 right-hand dummy zero

3

2 right-hand dummy zeros

4

Time display in seconds

5

Time display in HH.MM.SS format

6

Remote display (H, K, L commands)

7

Single-value remote display

8

Show 1

st

string value, slaved to another meter

9

Show 2

nd

string value, slaved to another meter

A

Show 3

rd

string value, slaved to another meter

B

Show 4

th

string value, slaved to another meter

C

Custom Start, Stop, Skip, Show

__0000

Counter mode

0

Basic counter

1

Extended counter

2

Extended counter, custom curve #1 selected

3

Extended counter, custom curve #2 selected

__0000

Linearization mode

0

Linear rate input.

1

Square root rate input.

__0000

Rate cutoff enable for
totalizing

0

Do not totalize rate values below

CutofFF

value.

(avoids totalizing small offsets from 0 rate value

or negative rate values).

1

Totalize all rates (required for bidirectional flow).

dSPyno

Display #

____01

PEAK key action

0

Display Peak

1

Display Valley

2

Peak (1

st

push), Valley (2

nd

push)

____01

Item to display after
Meter Reset*

1

Item #1*

2

Item #2*

3

Item #3*

CutofFF

Totalizing
cutoff*

_00000 _00000 _00000
_00000 _00000

Select digit to flash.

Select

0

thru

9

for flashing digits. In A_Atot or Batch

modes, meter will not totalize rate values below this
cutoff to avoid totalizing small offsets from zero.

GAtE_t

Gate time*

_000.00 _000.00 _000.00
_000.00 _000.00

Select digit to flash.

Select

0

thru

9

for flashing digit to set gate time* in

seconds. Decimal point location is fixed for 10 ms
resolution. This is time over which rate is measured.

.

ti_Out

.

Timeout*

_000.00 _000.00 _000.00
_000.00 _000.00

Select digit to flash.

Select

0

thru

9

for flashing digit to set timeout* in

seconds. Decimal point location is fixed for 10 ms
resolution. This is time during which batch relay is
de-energized at the end of a batch cycle.