beautypg.com

Cell constant selection, Range selection for 4 to 20 ma output, 10 cell – GF Signet 2850 Conductivity-Resistivity Sensor Electronics and Integral System User Manual

Page 5: Co o, 0 to 100 μs

background image

5

2850 Conductivity/Resistivity Sensor Electronics

• The Range selection switch bank (SW2) provides eight range

selections for each cell constant. Each range can be inverted,
making a total of 16 range options.

• Select a range from the table below and set SW2 as indicated.

O = Open

C = Closed

Example (refer to shaded selections of chart):
• The electrode installed is the 3-2840-1, with a 0.10 cell constant.
• The required output is 4 to 20 mA = 0 to 100 μS.
• Set SW2 switch bank to C-C-O-O.
• To invert range (4 to 20 mA = 100 to 0 μS), close switch #4.

1

2

3

4

ON

CTS

1

2 3

S

W1

SW3

SW4

D3

1

2

3

4

ON

CTS

1

2

3

4

ON

CTS

1

2

3

4

ON

CTS

1

2

3

4

ON

CTS

1

2

3

4

ON

CTS

1

2

3

4

ON

CTS

1

2

3

4

ON

CTS

1

2

3

4

ON

CTS

1

2

3

4

ON

CTS

Ch 1

Ch 2

1

2

3

4

ON

CTS

0.01/cm-1

0.1/cm-1

1.0/cm-1

10.0/cm-1

20.0/cm-1

Open

Closed

Open

Closed

Closed

Open

Closed

Open

Closed

Open

Closed

Open

Closed

Open

Example: 1.0 cell constant, temp comp ON

Example: 1.0 cell constant, temp comp OFF

Closed

Open

Custom Cell Constant

8. Cell Constant Selection

Cell Constant Selection
• The 2850 is shipped from the factory with the custom cell

constant and temperature offset programmed into the single-
input electronics. SW3 will be preset to "Custom cell constant".

• Switch banks SW3 and SW4 are used to select the cell

constant of the electrode. The Dual Input model is illustrated.
Single input models use only SW3.

• Make all switch settings before connecting power. Switch

changes made with the power ON will take 15 to 20 seconds
before becoming effective.

• Use SW3 #1-3 to select the cell constant for the fi rst sensor.
• Use SW3 #4 to disable the PT1000 Temp Compensation

function in the 2850 (as required for USP applications.) This
disables the function for BOTH INPUTS.

• Use SW4 #1-3 to select the cell constant for the second sensor.
• Set SW4 #4 to OPEN to disable the second input.

SW1

SW2

SW2

SW3

D3

1

2

3

4

ON

CTS

1

2

3

4

ON

CTS

1

3

2

1

2

3

4

ON

CTS

SW3 Cell Constant and Temp Compensation
#1 #2 #3 #4 For

Cell

Constant

C C C O Custom
C O O O 0.01

cm

-1

O C O O 0.1

cm

-1

C C O O 1.0

cm

-1

O O C O 10.0

cm

-1

C O C O 20.0

cm

-1

#4:

Open = Temp Comp ON

Closed = Temp Comp OFF

SW4: #1, #2 and #3 operate the same as SW3.

#4 turns input 2 OFF when set to OPEN.

(Open = Off, Closed = On)

9. Range Selection for 4 to 20 mA Output

SW2 Switch Setting

0.01 Cell

0.10 Cell

1.0 Cell

10.0 Cell

20.0 Cell

#1

#2

#3

#4

Resistivity Ranges in BOLD

C

C

C

O

10 to 20 MΩ

0 to 2 μS

0 to 20 μS

0 to 200 μS

0 to 400 μS

C

C

C

C

20 to 10 MΩ

2 to 0 μS

20 to 0 μS

200 to 0 μS

400 to 0 μS

O

C

C

O

2 to 10 MΩ

0 to 5 μS

0 to 50 μS

0 to 500 μS

0 to 1000 μS

O

C

C

C

10 to 2 MΩ

5 to 0 μS

50 to 0 μS

500 to 0 μS

1 000 to 0 μS

C

O

C

O

0 to 2 MΩ

0 to 10 μS

0 to 100 μS

0 to 1 000 μS

0 to 2 000 μS

C

O

C

C

2 to 0 MΩ

10 to 0 μS

100 to 0 μS

1 000 to 0 μS

2 000 to 0 μS

O

O

C

O

0 to 1 μS

0 to 50 μS

0 to 5 000 μS

0 to 5 000 μS

0 to 10 000 μS

O

O

C

C

1 to 0 μS

50 to 0 μS

500 to 0 μS

5 000 to 0 μS

10 000 to 0 μS

C

C

O

O

0 to 5 μS

0 to 100 μS

0 to 1 000 μS

0 to 10 000 μS

0 to 20 000 μS

C

C

O

C

5 to 0 μS

100 to 0 μS

1 000 to 0 μS

10 000 to 0 μS

20 000 to 0 μS

O

C

O

O

0 to 10 μS

0 to 200 μS

0 to 2 000 μS

0 to 50 000 μS

0 to 100 000 μS

O

C

O

C

10 to 0 μS

200 to 0 μS

2 000 to 0 μS

50 000 to 0 μS

100 000 to 0 μS

C

O

O

O

0 to 50 μS

0 to 500 μS

0 to 5 000 μS

0 to 100 000 μS

0 to 200 000 μS

C

O

O

C

50 to 0 μS

500 to 0 μS

5 000 to 0 μS

100 000 to 0 μS

200 000 to 0 μS

O

O

O

O

0 to 100 μS

0 to 1 000 μS

0 to 10 000 μS

0 to 200 000 μS

0 to 400 000 μS

O

O

O

C

100 to 0 μS

1000 to 0 μS

10 000 to 0 μS

200 000 to 0 μS

400 000 to 0 μS

C = CLOSED O = OPEN

Switch #4 inverts the output: OPEN = 4 to 20 mA, CLOSED = 20 to 4 mA

NOTE: To work correctly
with the 9900, the 2850 must
be set for the custom cell
constant or the actual probe
cell constant and the 9900
set for a 1.0 cell constant.