Signet 2750 DryLoc pH/ORP Sensor Electronics

9. Troubleshooting

LED and Output Condition

Possible Causes

Suggested Solutions

Current Out:
LED off, current output is 3.6 mA
S

3

L: Temp out is +999.9

• No electrode installed.
• Bad/dirty contacts between electrode

and 2750.

• Install

electrode.

• Check interconnection between

electrode and 2750, clean contacts.

During EasyCal, the LED blinks rapidly
for 4 seconds and the current output is
frozen at a random fi xed value.

• The buffer solution is outside of the

accepted tolerance for the 2750.

• The electrode is depleted (> 1.1 pH or

65 mV offset).

• pH system: Use fresh 4 pH, 7 pH,

or 10 pH buffer and restart the
calibration.

• ORP system: Use fresh 4pH and

7 pH buffer solution saturated with
quinhydrone.

• Replace the electrode.

After completing calibration procedure,
the output values are inaccurate.

• Insuffi cient time allowed for electrode

stabilization during calibration.

• Recalibrate; verify that test solutions

are at room temperature and wait
at least 30 seconds after placing
electrode in solution before pressing
S1 EasyCal button.