Signet 2750 DryLoc pH/ORP Sensor Electronics

Use standard installation
hardware to connect the
submersible 2750-3 or -4
directly to external equipment.

1/2 in. NPT threads

The 8052-1 NPT Mount Junction box
(or 8052-2 with EasyCal) connects to

3

/

4

in. pipe or conduit and provides

convenient wiring termination.

The 8050-1 Universal Mount junction
box (or 8050-2 with EasyCal) mounts
fl at onto a wall or can be strapped to
a post or pipe.

2750-3: 3/4 in. NPT threads
2750-4: ISO 7-1/R

3

/

4

threads

1.

Insert electrode into base

of the 2750-3 or 2750-4

and turn until keyed
contacts are seated.

2.

Thread retaining cap over

electrode and hand-tighten

onto

2750.

3.

Attach

3

/

4

in. watertight pipe to the top of the 2750.

Secure the threaded connection to prevent any leakage.

• For additional defense against possible accumulation of

condensation at the back seal area of the sensor, fi ll the
lower 3-4 inches (75-100 mm) of conduit or extension
pipe with a fl exible sealant such as silicone.

5. 2750-3, -4 Submersible Sensor Assembly and Installation

Note:
Keep the electrical interconnection between electrode and sensor
electronics dry and clean at all times.

94 mm

(3.7 in.)

107 mm

(4.2 in.)

Fill with

3 to 4 in.

of sealant

Fill with

3 to 4 in.

of sealant

106 mm

(4.2 in.)

94 mm

(3.7 in.)

OR

OR

Lubricate O-rings with a non-
petroleum based, viscous
lubricant (grease) compatible
with the system.

208 mm

(8.2 in.)

55 mm

(2.16 in.)

Fill with
75-100 mm
(3 to 4 in.)
of sealant

Cable supplied: 5 m (15 ft) 3-conductor + shld, 22 AWG.
May be extended up to 183 m (600 ft) with current output.
May be extended up to 305 m (1000 ft) with Digital
(S

3

L) output.