beautypg.com

Digital (s, L) wiring, 4 to 20 ma loop wiring – GF Signet 2750 DryLoc pH_ORP Sensor Electronics User Manual

Page 5

background image

5

Signet 2750 DryLoc pH/ORP Sensor Electronics

• Connect the 2750 cable directly to a Loop device as shown.
• This

confi guration does not provide any calibration capability

within the 2750 system. Periodic calibration must be
performed at the external equipment.

6. Digital

(S

3

L) wiring

• When the 2750 is powered with 5 VDC, the digital (S

3

L) serial data output is automatically selected.

• S

3

L data is used exclusively by Signet instruments.

• Remove approximately 10 mm (0.4 in.) of insulation and tin each conductor before inserting into connectors.

• Connect the 2750 cable directly to S

3

L I/O terminals.

8900 users: If this direct wiring is used, set the CALIBRATE

menu to "perform calibration at INSTRUMENT"

• When the 2750 includes a terminal block, connect the 2750

terminals as shown to any S

3

L I/O port.

8900 users: If the 2750 includes the EasyCal accessory,

set the CALIBRATE menu to "perform Calibration at either
SENSOR or INSTRUMENT"

• If SENSOR, use standard pH buffer values (pH 4, 7, or 10) to

perform periodic calibration.

• If INSTRUMENT, any pH value can be used.

• When the 2750 includes a terminal block, connect the 2750

terminals to the Loop device as shown.

• If the 2750 includes the EasyCal accessory, use standard pH

buffer values (pH 4, 7, or 10) to perform periodic calibration.

7. 4 to 20 mA Loop wiring

• When the 2750 is powered with 12 to 24 VDC, the 4 to 20 mA loop output is automatically selected.
• Remove approximately 10 mm (0.4 in.) of insulation and tin each conductor before inserting into connectors.

Current Loop with no junction box

Current loop with Junction box

S

3

L with no junction box

S

3

L with Junction box

Optional
Earth ground

1

2

3

4

+5 VDC

S

3

L data

Signal Ground

+

S

-

SW2

Run

Cal

SW1

D1

1

2

3

4

1

2

3

4

4 to 20 mA loop

monitor

Optional

Earth Ground

Power Supply

DC 12 to 24 V

+

-

+

S

-

SW2

Run

Cal

SW1

D1

1

2

3

4

4 20

4 to 20 mA loop

monitor

Optional
Earth Ground

Power Supply

DC 12 to 24 V

+

-

Black

Red (no connection)

White

Shield

4

20

+5 VDC

S

3

L data

Signal Ground

Black

Red

White

Shield