66

6)

Setting RBW

Command : RBW* * *

2650A, 2652A, 2658A :

( * * * * = 3K, 10K, 30K, 100K, 300K, 1M, 3M, AUTO, ALL [ unit : Hz ])

2651 :

( * * * * = 3K, 9K, 30K, 120K, 300K, 1M, 3M, AUTO, ALL [ unit : Hz ])

* ALL : The optimum RBW, VBW, and sweep time are set based on the frequency

span.

* AUTO : The optimum RBW is set based on the frequency span.

Example: RBW3M

Query: RBW?

7)

Setting VBW

Command : VBW* * * *

( * * * * = 100, 300, 1K, 3K, 10K, 30K, 100K, 300K, 1M, AUTO, ALL [ unit : Hz ] )

* ALL : The optimum RBW, VBW and sweep time are set based on the frequency

span.

* AUTO : The optimum RBW is set based on the frequency span.

Example: VBW300

Query: VBW?

8)

Selecting Measuring Function

Command : MEAS* * *

(* * * = CP, ACP, OBW, EF, MF,

FC, OFF)

Example: MEASOBW

(To change measuring function when in

a different measurement mode, set function OFF first by sending MEASOFF)

Query: MEAS?

9)

Request measuring result

Command : MEASRES

* Example of measuring result

Channel power measurement ----- POW: - 25.5 dBm

Adjacent channel power measurement ----- L: - 47.7 dBc U: - 48.3 dBc

Occupied bandwidth measurement ----- C: 1.45 G W: 20.00 k

Frequency counter ----- FC: 2400.0000 M

* If the frequency counter is not installed, “Invalid for F/C” is responded.

* If the level of spectrum is low and unmeasurable, “Non Signal” is

responded.

Command

Measuring function

CP

Channel power measurement

ACP

Adjacent channel power measurement

OBW

Occupied bandwidth measurement

EF

Electric field strength measurement

MF

Magnetic field strength measurement

FC

Frequency counter (factory option)

OFF

OFF