beautypg.com

6 setting of input impedance, 7 reference level setting range for each unit, Etting of – B&K Precision 2658A - Manual User Manual

Page 32: Nput, Mpedance, Eference, Evel, Etting, Ange for

background image

26

* If the unit is changed to

dBμV, dBmV, dBV or dBm, the offset is automatically changed.

2. By pushing

, the step size of offset is changed. (10 dB, 1 dB, 0.1 dB)

9.6 Setting of Input Impedance

By pushing , the input compensation can be set to 50

Ω (no offset) or 75 Ω. (5.6 dB offset

compensation)

The reference level is adjusted automatically, based on the input impedance selection. When the input impedance

is set to 75

Ω, the reference level is displayed including offset and conversion of 75Ω. For the amplitude reading to

be correct, a coaxial

50 Ω /75 Ω impedance conversion adapter must be connected to the RF input.

* When “75

Ω” is selected, values in Amplitude axis setting values display area will change according to

75

Ω, and the offset is set to 5.7 dB (insertion loss of 50 Ω /75 Ω adapter ). Moreover, the offset can be

changed.

When the unit at marker point is set to W, V, V/m or other, it is converted correctly from dBm.

* Be sure to attach a coaxial

50 Ω /75 Ω impedance conversion adapter when selecting “75 Ω”

9.7 Reference Level Setting Range for Each Unit

Unit

dBm

dBμV

dBmV

dBV

Maximum

10

117

57

-3

Minimum

-40

67

7

-53

Minimum

(shifted spectrum data)

-60

47

-13

-73

“Available unit in measuring function”

Unit

dBμV/m (Electric field strength measurement)

dBμA/m

(Magnetic field strength

measurement)

Setting

M401

M402

M403

M404

M405

M406

CP-2S

Maximum

143

146

149

150

137

159

160 ~ 203

Minimum

93

96

99

100

87

109

110 ~ 153

Minimum

(shifted spectrum

data)

73

76

79

80

67

89

90 ~ 133

* When the reference level is set between “Minimum” and “Minimum (shifted spectrum data)”, the

spectrum of “Minimum” is shifted and displayed on the screen.

F3

F6

This manual is related to the following products: