4 normalizing function, Ormalizing, Unction – B&K Precision 2658A - Manual User Manual

Page 63

background image

57

2. Each time

is pressed, T.G. output is alternately switched to ON or OFF.

22.4 Normalizing Function

NORM ON: The input level is made flat to the red line on the screen.

1. When is pushed, the same function menu as the previous item is displayed.

2. When

is pushed, TG. MODE is set.

3. Each time

is pushed, the normalizing function is alternately switched to ON or OFF.

If the normalizing function is switched to ON, “NORM ON” is displayed on the screen.

* If the setting is changed as follows, the normalizing function is automatically turned off.

· The span is expanded. · The center frequency is changed beyond the range normalized.

( When SPAN is changed from FULL SPAN according to center frequency, it is turned off.)

· The magnetic field strength measurement is selected.

· The AUTO tuning is executed. · The power supply is turned off. · The presetting is executed.

* When the scale is 2 dB or 5 dB, the normalizing function does not operate correctly if the spectrum level is not

displayed at a proper position on the screen.