beautypg.com

5 setting of trigger source, Auto tuning

Page 38: Etting of, Rigger, Ource

background image

32

By pushing , the detection mode is set.

POS SMPL NEG

POS (Positive Peak): The maximum value of the sample points is detected.

SMPL (Sample): The momentary value of the sample points is detected.

NEG (Negative Peak): The minimum value of the sample points is detected.

13.5 Setting of Trigger Source

By pushing , the trigger source is set.

INT EXT

INT: The sweep is automatically repeated. This setting is normally used.

EXT: When the signal over the trigger level (0.56 V) is input to external trigger input, the sweep

starts. The sweep automatically starts without an input. (Available only for zero span)

14. AUTO Tuning

When

is pushed, the spectrum with the maximum level is searched within full span, and then

it is adjusted to the center of the screen, and the optimum setting parameters are set.

* Set the frequency span before setting AUTO tuning.

* There is no function menu displayed for this mode. Auto tuning will execute upon pushing the key.

* The auto tuning does not operate normally in the following conditions.

1) Zero span

2) Full span

3) The signal level is –40 dBm or lower.

4) The signal frequency is 50 MHz or lower.

15. Hold / Run

By pushing , the sweep is switched to halt and continuance.

* There is no function menu displayed for this mode.

F4

F5

AUTO TUNE

HOLD / RUN

This manual is related to the following products: