19.3 Occupied Bandwidth Measurement

It is possible to measure the occupied frequency bandwidth defined as the width of points at N(%) of the total

power (N% POWER), or as the width of points that are X(dB) lower than the peak level (XdB DOWN).

N% POWER mode [ By pushing (MODE), N% POWER mode (N%) is selected.]

The bandwidth of points at N(%) of the total power is measured.

XdB DOWN mode

[ By pushing (MODE), XdB DOWN mode (X dB) is selected.]

The bandwidth of points that are X(dB) lower than the peak level is measured.

19.4 Electric Field Strength Measurement

Connecting a dipole antenna (option) to the RF input enables the measurement of the electric field strength.

Other than the optional antenna is also available by creating the original compensation data.

(Refer to “25.8 Writing of Original Compensation Data” for the details.)

F2

F3

[Measuring mode]

[Measured value]

[Setting parameter]

* “OBW XdB” is displayed in MEAS area on the screen.

* The measured value is displayed at the lower right corner on the screen.

1. By (XdB) , the down level from

the peak level is set.

* Setting range : 0.1 to 99.0dB

2. By , the step size of XdB can be set to 1dB or 0.1dB.

* “OBW N%” is displayed in MEAS area on the screen.

* The measured value is displayed at the lower right corner on the screen.

1. By (RATIO %) , the percentage of the

total power is set.

* Setting range : 80.0 to 99.9%

2. By , the step size of RATIO % can be set to 1% or 0.1%.

F2

F3

F3

F1

[Measuring mode]

[Setting parameter]

[Measured value]

F1

F4