beautypg.com

Center frequency

Page 27: Etting with step keys, Etting with encoder

background image

21

7. Center Frequency

When

is pushed, the following function menu is displayed.

7.1 Setting with step keys ( [F1], [F2] )

1. When is pushed, the center frequency decreases in the set step size.

2. When is pushed, the center frequency increases in the set step size.

3. Setting step size:

When is pushed, the step size is changed according to the following.

AUTO 100MHz 10MHz 1MHz 100kHz 20kHz


AUTO : The step size is 1/10 of the frequency span.

However, only for 500 kHz frequency span, the step size is set to 50 kHz.

If the frequency span is less than 500 kHz, the step size is set to 20 kHz.

AUTO is recommended for normal use.

7.2 Setting with encoder

1. By rotating , the center frequency is changed in the set step size.

2. Setting step size:

When is pushed, the step size is changed according to the following.

AUTO 100MHz 10MHz 1MHz 100kHz 20kHz


AUTO : The step size is 1/500 of the frequency span

However, if the frequency span is less than 5 MHz, the step size is set to 20 kHz.

AUTO is recommended for normal use.

CENTER FREQ

KeyStep

100M

EncStep

0.1M

SET

MKR

Ten Key

Mode

F1

F2

F3

F4

F5

F6

FREQ

* Center frequency setting range: 0 to 3.3 GHz for 2650A/2652A/2651, 0 to 8.5 GHz for 2658A

* The center frequency may shift for a while (up to 10 sec.), after setting is changed.

F1

F2

F3

F4

This manual is related to the following products: