beautypg.com

1 switching frequency band, Witching, Requency – B&K Precision 2658A - Manual User Manual

Page 30

background image

24

4. When is pushed and then the setting is FULL or ZERO span, the frequency span is returned to

the last setting.

5. By pushing ~ , the frequency span in function menu is set.

8.1 Switching Frequency Band

2658A has three frequency bands.

The frequency band is automatically set to the appropriate band based on the center frequency and span.

(At the span less than 200 MHz, only one band is used.)

When the setting frequency range belongs to two bands, the lower band has priority.

The frequency connection point of two bands is fixed as follows.

Two bands

Frequency connection point

Base band and Band 1

-

3.4 GHz

Band 1

-

and Band 1+

6.2 GHz

Note: The spectrum may change a little at the frequency connection point.

For accurate measurement, center frequency and span should be set as the measured frequency range in one band.

Frequency band

Measured frequency range

Base band

50 kHz to 3.5 GHz

Band 1

-

3.3 GHz to 6.3 GHz

Band 1+

6.1 GHz to 8.5 GHz

F1

F4

F6