B&K Precision 2658A - Manual User Manual
Page 56
50
“Specifications of magnetic field probe CP-2S” (Refer to the operating manual of CP-2S for the details.)
• Mode Selection and Measurement
By pushing (PROBE), CP-2S or USER mode is selected.
The measurement starts as soon as the probe is selected.
* “M/F CP2S” or “M/F USER” is displayed in MEAS area on the screen.
* “USER” is the original compensation data which is created by the user.
(Refer to “25.8 Writing of Original Compensation Data” for the details.)
The unit of amplitude axis automatically changes into [d
BμA/m].
* A spectrum other than the frequency range of the probe is not displayed.
Item
Specifications
Frequency range
10 MHz to 3 GHz
Space resolution (-6 dB)
approx. 0.25 mm
(depending on objects)
Reference level setting range: maximum
160 to 203
dBμA/m
Reference level setting range: minimum
(except for the minimum value in screen shift)
110 to 153 d
BμA/m
Measurement error
approx. ±1 dB (probe simple substance)
[Magnetic field probe connection]
The tip of the magnetic field probe CP-2S is made of glass-ceramic board. Take
care when handling the probe even though the strength of the glass-ceramic
board is durable under normal operation.
F1