19.6 Frequency Counter

This function currently not supported.

20. EMI Test (2651)

20.1 Additional Function for EMI Test

Detection mode (Refer to “13.4 Setting of Detection mode”)

By pushing

, the detection mode can be set.

POS SMPL NEG QP AV

In the 2651, QP detection and AV detection are added for EMI measurement.

POS (Positive Peak)

: The spectrum of the maximum value between the sample points

SMPL (Sample)

: The spectrum of the momentary value between the sample points

NEG (Negative Peak) : The spectrum of the minimum value between the sample points

QP (Quasi Peak)

: The spectrum of the quasi peak value between the sample points

AV (Average)

: The spectrum of the average value between the sample points


As for QP detection, the following characteristics are chosen according to RBW settings.

(based on CISPR16)

RBW

charging

time constant

dis-charging

time constant

meter

time constant

9kHz

1ms

160ms

160ms

120kHz

1ms

550ms

100ms

• Preset (Initialization) (Refer to “18. Save / Load”)

In the 2651, the radiated emission measurement and the conducted emission measurement are added to the

preset as a default setting.

By pushing

, the menu is displayed

NORMAL : The initial parameters of normal mode are set.

EMI-C : The initial parameters of conducted emission measurement are set.

EMI-R : The initial parameters of radiated emission measurement are set

In the radiated mission measurement, “USER” is selected as an antenna.

It is necessary to input the compensation data of antenna beforehand.

F6

F6

SWEEP

F6

SAVE/LOAD

F1

F2

F3