Table 6. Corrected system performance with Type-N 50 Ω connectors,
85032F calibration kit, enhanced response calibration
Network analyzer: E5061B-115/135/215/235, calibration kit: 85032F (Type-N, 50 Ω),
calibration: enhanced response calibration
IF bandwidth = 10 Hz, No averaging applied to data, environmental temperature
= 23 °C ±5 °C with < 1 °C deviation from calibration temperature, isolation calibration
not omitted
Description
Specification (dB)
100 kHz to 300 kHz
300 kHz to 1 MHz
1 MHz to 3 GHz
Directivity
49
49
46
Source match
41
41
40
Load match
48
49
46
Refl ection tracking
0.011
0.011
0.021
Transmission tracking
0.035
0.028
0.034
0.01
0.1
1
10
-90
-80
-70
-60
-50
-40
-30
-20
-10
10
0
Uncertainty (dB)
Transmission Coefficient (dB)
Magnitude
100 kHz to 300 kHz
300 kHz to 1 MHz
1 MHz to 3 GHz
S11 = S22 = 0; Source Power = -10 dBm
0
0.01
0.02
0.03
0.04
0.05
0
0.2
0.4
0.6
0.8
1
U
n
cer
tai
n
ty (
Li
n
ear
)
Reflection Coefficient (Linear)
Magnitude
100 kHz to 300 kHz
300 kHz to 1 MHz
1 MHz to 3 GHz
S11 = S22 = 0; Source Power = -10 dBm
0
2
4
6
8
10
0
0.2
0.4
0.6
0.8
1
Uncertainty (Degrees)
Reflection Coefficient (Linear)
Phase
100 kHz to 300 kHz
300 kHz to 1 MHz
1 MHz to 3 GHz
S11 = S22 = 0; Source Power = -10 dBm
0.1
1
10
100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Uncertainty (Degrees)
Transmission Coefficient (dB)
Phase
100 kHz to 300 kHz
300 kHz to 1 MHz
1 MHz to 3 GHz
S11 = S22 = 0; Source Power = -10 dBm
Figure 9. Transmission uncertainty (specifi cation)
Figure 10. Refl ection uncertainty (specifi cation)