beautypg.com

Atec Agilent-E5061B User Manual

Page 21

background image

21

Table 26. Test port input (Trace noise)

Description

Specification

Typical

Trace noise

(at IF automatic bandwidth,
<10 kHz)

5 mdB rms

(at 3 kHz bandwidth, 10 kHz to
30 MHz at -5 dBm, 0 dB
attenuation, 50 Ω)

5 mdB rms

Trace noise phase

(at IF automatic bandwidth,
<10 kHz)

0.03° rms

(at 3 kHz bandwidth, 10 kHz to
30 MHz at -5 dBm, 0 dB
attenuation, 50 Ω)

0.03° rms

Table 27. Test port input (stability)

Description

Specification

SPD

Stability magnitude

< ±0.02 dB/°C

Stability phase

< ±0.2°/°C

Table 28. Test port input (Dynamic Accuracy)

1

Description

Specification

Typical

Dynamic accuracy magnitude
(0 dB attenuation, 50 Ω)

±0.303 dB at -5 dBm
±0.09 dB at -15 dBm
±0.056 dB at -25 dBm
±0.073 dB at -35 dBm
±0.087 dB at -45 dBm
±0.103 dB at -55 dBm
±0.121 dB at -65 dBm
±0.15 dB at -75 dBm
±0.211 dB at -85 dBm
±0.371 dB at -95 dBm
±0.841 dB at -105 dBm
±2.141 dB at -115 dBm

Dynamic accuracy phase
(0 dB attenuation, 50 Ω)

±2.04 ° at -5 dBm
±0.6 ° at -15 dBm
±0.37 ° at -25 dBm
±0.48 ° at -35 dBm
±0.58 ° at -45 dBm
±0.68 ° at -55 dBm
±0.81 ° at -65 dBm
±1.00 ° at -75 dBm
±1.41 ° at -85 dBm
±2.5 ° at -95 dBm
±5.83 ° at -105 dBm
±16.23 ° at -115 dBm

±5 °
(+15 dBm, 20
dB attenuation)

1. Accuracy of the test port input power reading is relative to –25 dBm reference input power level.