beautypg.com

Description specification (db) – Atec Agilent-E5061B User Manual

Page 10

background image

10

Table 8. Corrected system performance with Type-N 75 Ω connectors,
85036B calibration kit, full 2-port calibration

Network analyzer: E5061B-117/137/217/237, calibration kit: 85036B (Type-N, 75 Ω),
calibration: Full 2 port

IF bandwidth = 10 Hz, No averaging applied to data, environmental temperature
= 23 °C ±5 °C with < 1 °C deviation from calibration temperature, isolation calibration
not omitted

Description

Specification (dB)

100 kHz to 300 kHz

300 kHz to 1 MHz

1 MHz to 3 GHz

Directivity

49

48

44

Source match

48

41

35

Load match

48

48

44

Refl ection tracking

0.004

0.010

0.019

Transmission tracking

0.022

0.028

0.052

0.01

0.1

1

10

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Uncertainty (dB)

Transmission Coefficient (dB)

Magnitude

100 kHz to 300 kHz

300 kHz to 1 MHz

1 MHz to 3 GHz

S11 = S22 = 0; Source Power = -10 dBm

Uncertainty (Linear)

Reflection Coefficient (Linear)

Magnitude

0

0.01

0.02

0.03

0.04

0.05

0

0.2

0.4

0.6

0.8

1

100 kHz to 300 kHz

300 kHz to 1 MHz

1 MHz to 3 GHz

S11 = S22 = 0; Source Power = -10 dBm

0

2

4

6

8

10

0

0.2

0.4

0.6

0.8

1

Uncertainty (Degrees)

Reflection Coefficient (Linear)

Phase

100 kHz to 300 kHz

300 kHz to 1 MHz

1 MHz to 3 GHz

S11 = S22 = 0; Source Power = -10 dBm

0.1

1

10

100

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Uncertainty (Degrees)

Transmission Coefficient (dB)

Phase

100 kHz to 300 kHz

300 kHz to 1 MHz

1 MHz to 3 GHz

S11 = S22 = 0; Source Power = -10 dBm

Figure 13. Transmission uncertainty (specifi cation)

Figure 14. Refl ection uncertainty (specifi cation)