Description specification (db) – Atec Agilent-E5061B User Manual
Page 10
10
Table 8. Corrected system performance with Type-N 75 Ω connectors,
85036B calibration kit, full 2-port calibration
Network analyzer: E5061B-117/137/217/237, calibration kit: 85036B (Type-N, 75 Ω),
calibration: Full 2 port
IF bandwidth = 10 Hz, No averaging applied to data, environmental temperature
= 23 °C ±5 °C with < 1 °C deviation from calibration temperature, isolation calibration
not omitted
Description
Specification (dB)
100 kHz to 300 kHz
300 kHz to 1 MHz
1 MHz to 3 GHz
Directivity
49
48
44
Source match
48
41
35
Load match
48
48
44
Refl ection tracking
0.004
0.010
0.019
Transmission tracking
0.022
0.028
0.052
0.01
0.1
1
10
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Uncertainty (dB)
Transmission Coefficient (dB)
Magnitude
100 kHz to 300 kHz
300 kHz to 1 MHz
1 MHz to 3 GHz
S11 = S22 = 0; Source Power = -10 dBm
Uncertainty (Linear)
Reflection Coefficient (Linear)
Magnitude
0
0.01
0.02
0.03
0.04
0.05
0
0.2
0.4
0.6
0.8
1
100 kHz to 300 kHz
300 kHz to 1 MHz
1 MHz to 3 GHz
S11 = S22 = 0; Source Power = -10 dBm
0
2
4
6
8
10
0
0.2
0.4
0.6
0.8
1
Uncertainty (Degrees)
Reflection Coefficient (Linear)
Phase
100 kHz to 300 kHz
300 kHz to 1 MHz
1 MHz to 3 GHz
S11 = S22 = 0; Source Power = -10 dBm
0.1
1
10
100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Uncertainty (Degrees)
Transmission Coefficient (dB)
Phase
100 kHz to 300 kHz
300 kHz to 1 MHz
1 MHz to 3 GHz
S11 = S22 = 0; Source Power = -10 dBm
Figure 13. Transmission uncertainty (specifi cation)