beautypg.com

Gain phase measurement (option 3l5 only) – Atec Agilent-E5061B User Manual

Page 19

background image

19

Gain Phase Measurement (Option 3L5 only)

Source Characteristics (LF Out)

Table 20. Source characteristics output frequency

Description

Specification

Typical

Range

5 Hz to 30 MHz

Resolution

1 mHz

Source stability

±7 ppm (5 °C to 40 °C)

CW accuracy

±7 ppm ±1 mHz

High stability option (Option 1E5)

CW accuracy

±1 ppm ±1 mHz

Source stability

±0.05 ppm (5 °C to 40 °C)

±0.5 ppm per year

Table 21. Source characteristics output power

Description

Specification

SPD

Level accuracy

±1 dB
(at 0 dBm absolute 200 Hz)

±2 dB
(at 0 dBm, relative to 200 Hz)

Level

±1 dB
(at -10 dBm to 10 dBm, 0
dBm reference)

±1 dB
(at -45 dBm to -10 dBm, 0
dBm reference)

Range

–45 dBm to 10 dBm

Sweep range

–45 dBm to 10 dBm

Level resolution

0.05 dB

Table 22. Source characteristics output signal purity

Description

Specification

Typical

Harmonics (2nd or 3rd)

< –20 dBc
(at 5 dBm)

Non-harmonic spurious

< –25 dBc
(at 5 dBm)

Table 23. Source characteristics output impedance

Description

Specification

Typical

Impedance

50 Ω nominal

Return loss

> 10 dBc