Table 2. Corrected system performance with Type-N 50 Ω connectors,
85032F calibration kit, full 2-port calibration

Network analyzer: E5061B-3L5, calibration kit: 85032F (Type-N, 50 Ω),
calibration: full 2-port

IF bandwidth = 10 Hz, No averaging applied to data, environmental temperature
= 23 °C ±5 °C with < 1 °C deviation from calibration temperature, isolation calibration
not omitted

Description

Specification (dB)

100 Hz to 100 kHz

100 kHz to 1 GHz

1 GHz to 3 GHz

Directivity

49

49

46

Source match

41

41

40

Load match

49

49

46

Refl ection tracking

0.011

0.011

0.021

Transmission tracking

0.019

0.019

0.026

0.01

0.1

1

10

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

100 Hz to 100 kHz
100 kHz to 1 GHz
1 GHz to 3 GHz

S11 = S22 = 0 ; Source power = -10 dBm

Magnitude

Transmission Coefficient (dB)

U

n

cer

tai

n

ty (

d

B

)

0

0.01

0.02

0.03

0.04

0.05

0

0.2

0.4

0.6

0.8

1

100 Hz to 100 kHz
100 kHz to 1 GHz
1 GHz to 3 GHz

S21 = S12 = 0 ; Source power = -10 dBm

Reflection Coefficient (Linear)

Magnitude

U

n

c

e

rt

a

in

ty

(

Linear)

0.1

1

10

100

100 Hz to 100 kHz
100 kHz to 1 GHz
1 GHz to 3 GHz

S11 = S22 = 0 ; Source power = -10 dBm

-9 0

-8 0

-7 0

-6 0

-5 0

-4 0

-3 0

-2 0

-1 0

0

1 0

Transmission Coefficient (dB)

Phase

U

n

cer

tai

n

ty (

D

eg

rees)

0

2

4

6

8

1 0

0

0 .2

0 .4

0 .6

0 .8

1

100 Hz to 100 kHz
100 kHz to 1 GHz
1 GHz to 3 GHz

S21 = S12 = 0 ; Source Power = -10 dBm

Reflection Coefficient (Linear)

Phase

U

n

c

e

rt

a

in

ty

(

Degrees)

Figure 1. Transmission uncertainty (specifi cation)

Figure 2. Refl ection uncertainty (specifi cation)