beautypg.com

Storage automation – Atec Agilent-E5061B User Manual

Page 34

background image

34

Storage

Automation

Internal hard disk drive

Store and recall instrument states, calibration
data, and trace data into internal hard drive. Trace
data can be saved in CSV (comma separated
value) format. All fi les are MS-DOS

®

-compatible.

Instrument states include all control settings,
limit lines, segment sweep tables, and memory
trace data.

File sharing

Internal hard disk drive (D:) can be accessed from
an external Windows

®

PC through LAN.

Screen hardcopy

Printouts of instrument data are directly produced
on a printer through USB interfaces.

System capabilities

Familiar graphical user interface

The analyzer employs a graphical user interface
based on Windows

®

operating system. There are

three ways to operate the instrument manually:
you can use a hardkey interface, touch screen
interface or a mouse interface.

Limit lines

Defi ne the test limit lines that appear on the
display for pass/fail testing. Defi ned limits may be
any combination of horizontal/sloping lines and
discrete data points.

GPIB/LAN/USB

Internal

SCPI

Ч

Ч

COM

×

Methods

Internal analyzer execution

Applications can be developed in a built-in
VBA

®

(Visual Basic for Applications) language.

Applications can be executed from within the
analyzer via COM (component object model) or
using SCPI.

Controlling via GPIB

The GPIB interface operates to IEEE 488.2 and
SCPI protocols. The analyzer can be controlled
by a GPIB external controller. The analyzer can
control external devices using a USB/GPIB
interface.

LAN

Protocol

TCP/IP

Function

Telnet, SICL-LAN

USB

Protocol

USB Test and Measurement Class (TMC)
interface that communicates over USB, complying
with the IEEE 488.1 and IEEE 488.2 standards.