Description specification (db) – Atec Agilent-E5061B User Manual
Page 5
5
Table 3. Corrected system performance with Type-N 50 Ω connectors,
85032F calibration kit, enhanced response calibration
Network analyzer: E5061B-3L5, calibration kit: 85032F (Type-N, 50 Ω),
calibration: enhanced response
IF bandwidth = 10 Hz, No averaging applied to data, environmental temperature
= 23 °C ±5 °C with < 1 °C deviation from calibration temperature, isolation calibration
not omitted
Description
Specification (dB)
100 Hz to 100 kHz
100 kHz to 1 GHz
1 GHz to 3 GHz
Directivity
49
49
46
Source match
41
41
40
Load match
49
49
46
Refl ection tracking
0.011
0.011
0.021
Transmission tracking
0.019
0.019
0.033
0.01
0.1
1
10
100 Hz to 100 kHz
100 kHz to 1GHz
1 GHz to 3 GHz
S11 = S22 = 0 ; Source power = -10 dBm
-9 0
-8 0
-7 0
-6 0
-5 0
-4 0
-3 0
-2 0
-1 0
0
1 0
Transmission Coefficient (dB)
Magnitude
U
n
cer
tai
n
ty (
d
B
)
0
0.01
0.02
0.03
0.04
0.05
0
0.2
0.4
0.6
0.8
1
100 Hz to 100 kHz
100 kHz to 1GHz
1 GHz to 3 GHz
S21 = S12 = 0 ; Source power = -10 dBm
Reflection Coefficient (Linear)
Magnitude
U
n
c
e
rt
a
in
ty
(
Linear)
0.1
1
10
100
100 Hz to 100 kHz
100 kHz to 1GHz
1 GHz to 3 GHz
S11 = S22 = 0 ; Source power = -10 dBm
Transmission Coefficient (dB)
Phase
U
n
c
e
rt
a
in
ty
(
Degrees)
-9 0
-8 0
-7 0
-6 0
-5 0
-4 0
-3 0
-2 0
-1 0
0
1 0
0
2
4
6
8
10
0
0.2
0.4
0.6
0.8
1
100 Hz to 100 kHz
100 kHz to 1GHz
1 GHz to 3 GHz
S21 = S12 = 0 ; Source Power = -10 dBm
Reflection Coefficient (Linear)
Phase
U
n
c
e
rt
a
in
ty
(
Degrees)
Figure 3. Transmission uncertainty (specifi cation)
Figure 4. Refl ection uncertainty (specifi cation)