beautypg.com

Measurement capabilities – Atec Agilent-E5061B User Manual

Page 31

background image

31

Measurement capabilities

Number of measurement channels

Up to 4 independent measurement channels. A
measurement channel is coupled to stimulus
response settings including frequency, IF
bandwidth, power level, and number of points.

Number of display windows

Each measurement channel has a display
window. Up to 4 display windows (channels) can
be displayed.

Number of traces

4 data traces and 4 memory traces per channel

Measurement choices

Option 115,135,117 and 137 - S11, S21, Absolute
Option 215, 235 and 237 - S11, S21, S12, S22, Absolute
Option 3L5 - S11, S21, S12, S22, T/R, T, R, Absolute

Measurement parameter
conversion

Available to convert S-parameters into refl ection
impedance, transmission impedance, refl ection
admittance, transmission admittance, and 1/S.

Data formats

Log magnitude, linear magnitude, phase,
expanded phase, positive phase, group delay,
SWR, real, imaginary, Smith chart, polar.

Data markers

10 independent markers per trace. Reference
marker available for delta marker operation. Smith
chart format includes 5 marker formats: linear
magnitude/phase, log magnitude/phase, real/
imaginary, R + jX, and G + jB. Polar chart format
includes 3 marker formats: linear magnitude/
phase, log magnitude/phase, and real/imaginary.

Marker functions

Marker search

Max value, min value, multi-peak, multi-target,
peak, peak left, peak right, target, target left,
target right, and width parameters with user-
defi ned bandwidth values.

Marker-to functions

Set start, stop, center to active marker stimulus
value; set reference to active marker response
value; set electrical delay to group delay at active
marker.

Search range

User defi nable.

Tracking

Performs marker search continuously or on
demand.

Fault location functions (Option 010)

Transformation to distance and
time domain

Selectable transformation type from bandpass,
lowpass impulse, lowpass step.
Selectable window from maximum, normal and
minimum.

Impedance Measurement Analysis (Option 005)

Impedance Measurement

Selectable Impedance Parameter and Equivalent
Circuit analysis Capability