Description general characteristics – Atec Agilent-E5061B User Manual
Page 26

26
Table 35. Analyzer environment and dimensions
Description
General Characteristics
Operating environment
Temperature
+5 °C to +40 °C
Error-corrected temperature range
23 °C ±5 °C with < 1 °C deviation from calibration
temperature
Humidity
20% to 80% at wet bulb temperature < +29 °C
(non-condensing)
Altitude
0 to 2,000 m (0 to 6,561 feet)
Vibration
0.21 G maximum, 5 Hz to 500 Hz
Non-operating storage environment
Temperature
–10 °C to +60 °C
Humidity
20% to 90% at wet bulb temperature < 40 °C
(non-condensing)
Altitude
0 to 4,572 m (0 to 15,000 feet)
Vibration
0.5 G maximum, 5 Hz to 500 Hz
Dimensions
Height = 235 mm, Width = 432 mm, Depth = 296 mm.
See fi gures 10 to 12.
Weight
14.0 kg
Magnetic Susceptibility
Degradation of some product specifi cations can
occur in the presence of ambient power frequency
magnetic fi elds of 30 A/m or greater.
The product self-recovers and operates as speci-
fi ed when removed or shielded from the ambient
magnetic fi eld.
When the analyzer tuned frequency is identical
to the immunity test signal frequency, there may
be signals of up to −80 dB of full-scale response
displayed on the screen.
Magnetic Emission
Emission of magnetic fi eld may occur at the left
side of the where two cooling fans are installed.
Its magnitude can be as much as 160A/m and
25A/m at 0cm and 1cm apart from the center of
the fan, respectively. It is recommended to have
enough clearance between the cooling fans and
magnetically sensitive device or instruments.