Description specification (db) – Atec Agilent-E5061B User Manual
Page 7
7
Table 5. Corrected system performance with Type-N 50 Ω connectors,
85032F calibration kit, full 2-port calibration
Network analyzer: E5061B-115/135/215/235, calibration kit: 85032F (Type-N, 50 Ω),
calibration: Full 2 port
IF bandwidth = 10 Hz, No averaging applied to data, environmental temperature
= 23 °C ±5 °C with < 1 °C deviation from calibration temperature, isolation calibration
not omitted
Description
Specification (dB)
100 kHz to 300 kHz
300 kHz to 1 MHz
1 MHz to 3 GHz
Directivity
49
49
46
Source match
41
41
40
Load match
48
49
46
Refl ection tracking
0.011
0.011
0.021
Transmission tracking
0.035
0.028
0.034
U
n
cer
tai
n
ty (
d
B
)
Transmission Coefficient (dB)
Magnitude
0 .0 1
0 .1
1
1 0
-9 0
-8 0
-7 0
-6 0
-5 0
-4 0
-3 0
-2 0
-1 0
0
1 0
100 kHz to 300 kHz
300 kHz to 1 MHz
1 MHz to 3 GHz
S11 = S22 = 0; Source Power = -10 dBm
0
0.01
0.02
0.03
0.04
0.05
0
0.2
0.4
0.6
0.8
1
U
n
cer
tai
n
ty (
Li
n
ear
)
Reflection Coefficient (Linear)
Magnitude
100 kHz to 300 kHz
300 kHz to 1 MHz
1 MHz to 3 GHz
S11 = S22 = 0; Source Power = -10 dBm
0
2
4
6
8
10
0
0.2
0.4
0.6
0.8
1
Uncertainty (Degrees)
Reflection Coefficient (Linear)
Phase
100 kHz to 300 kHz
300 kHz to 1 MHz
1 MHz to 3 GHz
S11 = S22 = 0; Source Power = -10 dBm
0.1
1
10
100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Uncertainty (Degrees)
Transmission Coefficient (dB)
Phase
S11 = S22 = 0; Source Power = -10 dBm
100 kHz to 300 kHz
300 kHz to 1 MHz
1 MHz to 3 GHz
Figure 7. Transmission uncertainty (specifi cation)