Source control trace functions – Atec Agilent-E5061B User Manual

Page 32

background image

32

Source control

Trace functions

Measured number of points per
sweep

User defi nable from 2 to 1601.

Sweep type

Linear sweep, segment sweep, log sweep , power
sweep and DC bias sweep.

Segment sweep

Defi ne independent sweep segments. Set number
of points, test port power levels, IF bandwidth,
delay time, sweep time independently for each
segment.

Sweep trigger

Set to continuous, hold, or single, sweep with
internal, external, manual, or bus trigger.

Power

Set source power from -45 dBm to 10 dBm. The
power slope function compensates source power
error.

Display data

Display current measurement data, memory
data, or current measurement and memory data
simultaneously.

Trace math

Vector addition, subtraction, multiplication or
division of measured complex values and memory
data.

Title

Add custom title to each channel window.
Titles are printed on hardcopies of displayed
measurements

Autoscale

Automatically selects scale resolution and
reference value to vertically center the trace.

Electrical delay

Offset measured phase or group delay by a
defi ned amount of electrical delay, in seconds.

Phase offset

Offset measured phase or group delay by a
defi ned amount in degrees.

Statistics

Calculates and displays mean, standard deviation
and peak-to-peak deviation of the data trace.