6
Table 4. Corrected system performance with Type-N 50 Ω connectors,
85092C ECal Module, full 2-port calibration
Network analyzer: E5061B-3L5, calibration kit: 85092C (Type-N, 50 Ω 300 kHz to 9 GHz),
calibration: Full 2 port
IF bandwidth = 10 Hz, No averaging applied to data, environmental temperature
= 23 °C ±5 °C with < 1 °C deviation from calibration temperature, isolation calibration
not omitted
Description
Specification (dB)
300 kHz to 10 MHz
10 MHz to 1 GHz
1 GHz to 3 GHz
Directivity
45
52
54
Source match
36
45
44
Load match
41
47
47
Refl ection tracking
0.100
0.040
0.040
Transmission tracking
0.056
0.039
0.040
Transmission Coefficient (dB)
Magnitude
U
n
cer
tai
n
ty (
d
B
)
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
0.01
0.1
1
10
1 GHz to 3 GHz
10 MHz to 1 GHz
300 kHz to 10 MHz
S11 = S22 = 0; Source Power = -10 dBm
300 kHz to 10 MHz
10 MHz to 1 GHz
1 GHz to 3 GHz
0
0.01
0.02
0.03
0.04
0.05
S21 = S12 = 0; Source Power = -10 dBm
Reflection Coefficient (Linear)
Magnitude
U
n
c
e
rt
a
in
ty
(
Linear)
0
0.2
0.4
0.6
0.8
1
300 kHz to 10 MHz
10 MHz to 1 GHz
1 GHz to 3 GHz
0.1
1
10
100
S11 = S22 = 0; Source Power = -10 dBm
Transmission Coefficient (dB)
Phase
Uncertainty (Degrees)
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
300 kHz to 10 MHz
10 MHz to 1 GHz
1 GHz to 3 GHz
S21 = S12 = 0; Source Power = -10 dBm
0
2
4
6
8
10
0
0.2
0.4
0.6
0.8
1
Reflection Coefficient (Linear)
Phase
U
n
c
e
rt
a
in
ty
(
Degrees)
Figure 5. Transmission uncertainty (specifi cation)