beautypg.com

Relative accuracy – Atec Agilent-E4980A User Manual

Page 9

background image

9

Relative accuracy

Relative accuracy includes stability, temperature coefficient, linearity, repeatability, and
calibration interpolation error. Relative accuracy is specified when all of the following
conditions are satisfied:

• Warm-up time: 30 minutes
• Test cable length: 0 m, 1 m, 2 m, or 4 m (Agilent 16047A/B/D/E)
• A “Signal Source Overload” warning does not appear.
When the test signal current exceeds a value in table 11 below, a “Signal

Source Overload” warning appears.

Table 11.

Test signal voltage

Test frequency

Condition

1

≤ 2 Vrms

> 2 Vrms

≤ 1 MHz

the smaller value of either 110 mA or

130 mA - 0.0015 × Vac × (Fm / 1 MHz) × (L_cable + 0.5)

> 1 MHz

70 mA - 0.0015 × Vac × (Fm / 1 MHz) × (L_cable + 0.5)

Vac [V]

Test signal voltage

Fm [Hz]

Test frequency

L_cable [m]

Cable length

• OPEN and SHORT corrections have been performed.
• Bias current isolation: Off
• The DC bias current does not exceed a set value within each range of the DC
bias

current

• The optimum impedance range is selected by matching the impedance of DUT to the
effective measuring range.
• Under an AC magnetic field, the following equation is applied to the measurement
accuracy.
A x ( 1 + B x ( 2 + 0.5 / Vs))
Where
A: Absolute accuracy
B: Magnetic flux density [Gauss]
Vs: Test signal voltage level [Volts]

|Z|, |Y|, L, C, R, X, G, and B accuracy (L, C, X, and B accuracies apply when
Dx ≤ 0.1, R and G accuracies apply Qx ≤ 0.1)

Relative accuracy Ae is given as:

Equation 8.

Ae = [Ab + Zs /|Zm| × 100 + Yo × |Zm| Ч 100 ] Ч Kt

Zm Impedance

of

DUT

Ab Basic

accuracy

Zs Short

offset

Yo Open

offset

Kt Temperature

coefficient

D accuracy

D accuracy De is given as
• when Dx ≤ 0.1

Equation 9.

De = ±Ae/100

Dx Measured

D

value

Ae

Relative accuracies of |Z|, |Y|, L, C, R, X, G, and B


• when Dx > 0.1, multiply De by (1 + Dx)

1. When the calculation result is a negative value,

0 A is applied.