beautypg.com

Atec Agilent-E4980A User Manual

Page 10

background image

10

Q accuracy (when Q x De < 1)

Q accuracy Qe is given as:

Equation 10.

Qe = ±

(Qx

2

× De)

(1 Qx × De)

Qx Measured

Q

value

De

Relative D accuracy

θ accuracy

θ accuracy θe is given as:

Equation 11.

θe = 180 × Ae (deg)

π × 100

Ae

Relative accuracies of |Z|, |Y|, L, C, R, X, G, and B

G accuracy (when Dx ≤ 0.1)

G accuracy Ge is given as:

Equation 12.

Ge = Bx × De

(S)

Bx = 2πfCx =

1

2πfLx

Ge

Relative G accuracy

Dx Measured

D

value

Bx Measured

B

value

De

Relative D accuracy

f

Test frequency

Cx

Measured C value (F)

Lx

Measured L value (H)

Rp accuracy (when Dx ≤ 0.1)

Rp accuracy Rpe is given as:

Equation 13.

Rpe = ±

Rpx × De

(Ω)

Dx De

Rpe Relative

Rp

accuracy

Rpx

Measured Rp value (Ω)

Dx Measured

D

value

De

Relative D accuracy

Rs accuracy (when Dx ≤ 0.1)

Rs accuracy Rse is given as:

Equation 14.

Rse = Xx × De

(Ω)

Xx = 1 = 2πfLx

2πfCx

Rse Relative

Rs

accuracy

Dx Measured

D

value

Xx

Measured X value (Ω)

De

Relative D accuracy

f

Test frequency (Hz)

Cx

Measured C value (F)

Lx

Measured L value (H)

±

±