Basic accuracy – Atec Agilent-E4980A User Manual

Page 11

background image

11

Example of C-D accuracy calculation

Measurement conditions

Test Frequency:

1

kHz

Measured C value:

100 nF

Test signal voltage: 1

Vrms

Measurement time mode:

Medium

Measurement temperature: 23 °C

Ab = 0.05%
|Zm| = 1 / (2π Ч 1 Ч 10

3

Ч 100 Ч 10

-9

) = 1590 Ω

Zs = 0.6 mΩ Ч (1 + 0.400/1) Ч (1 + √(1000/1000) = 1.68 mΩ
Yo = 0.5 nS Ч (1 + 0.100/1) Ч (1 + √(100/1000) = 0.72 nS
C accuracy: Ae = [0.05 + 1.68 m/1590 × 100 + 0.72 n Ч 1590 Ч 100] Ч 1 = 0.05%
D accuracy: De = 0.05/100 = 0.0005

Basic accuracy

Basic accuracy Ab is given below.

Table 12. Measurement time mode = SHORT

Test signal voltage

Test

5 mVrms -

50 mVrms -

0.3 Vrms -

1 Vrms -

10 Vrms -

frequency [Hz] 50 mVrms

0.3 Vrms

1 Vrms

10 Vrms

20 Vrms

20 - 125

(0.6%) ×

0.60%

0.30%

0.30%

0.30%

(50

mVrms/Vs)

125 - 1 M

(0.2%) ×

0.20%

0.10%

0.15%

0.15%

(50

mVrms/Vs)

1 M - 2 M

(0.4%) ×

0.40%

0.20%

0.30%

0.30%

(50 mVrms/Vs)

Table 13. Measurement time mode = MED, LONG

Test signal voltage

Test

5 mVrms -

50 mVrms -

0.3 Vrms -

1 Vrms -

10 Vrms -

frequency [Hz] 50 mVrms

0.3 Vrms

1 Vrms

10 Vrms

20 Vrms

20 - 100

(0.25%) ×

0.25%

0.10%

0.15%

0.15%

(30

mVrms/Vs)

100 - 1 M

(0.1%) ×

0.10%

0.05%

0.10%

0.15%

(30

mVrms/Vs)

1 M - 2 M

(0.2%) ×

0.20%

0.10%

0.20%

0.30%

(30

mVrms/Vs)

Vs [Vrms]

Test signal voltage