Basic accuracy – Atec Agilent-E4980A User Manual
Page 11
11
Example of C-D accuracy calculation
Measurement conditions
Test Frequency:
1
kHz
Measured C value:
100 nF
Test signal voltage: 1
Vrms
Measurement time mode:
Medium
Measurement temperature: 23 °C
Ab = 0.05%
|Zm| = 1 / (2π Ч 1 Ч 10
3
Ч 100 Ч 10
-9
) = 1590 Ω
Zs = 0.6 mΩ Ч (1 + 0.400/1) Ч (1 + √(1000/1000) = 1.68 mΩ
Yo = 0.5 nS Ч (1 + 0.100/1) Ч (1 + √(100/1000) = 0.72 nS
C accuracy: Ae = [0.05 + 1.68 m/1590 × 100 + 0.72 n Ч 1590 Ч 100] Ч 1 = 0.05%
D accuracy: De = 0.05/100 = 0.0005
Basic accuracy
Basic accuracy Ab is given below.
Table 12. Measurement time mode = SHORT
Test signal voltage
Test
5 mVrms -
50 mVrms -
0.3 Vrms -
1 Vrms -
10 Vrms -
frequency [Hz] 50 mVrms
0.3 Vrms
1 Vrms
10 Vrms
20 Vrms
20 - 125
(0.6%) ×
0.60%
0.30%
0.30%
0.30%
(50
mVrms/Vs)
125 - 1 M
(0.2%) ×
0.20%
0.10%
0.15%
0.15%
(50
mVrms/Vs)
1 M - 2 M
(0.4%) ×
0.40%
0.20%
0.30%
0.30%
(50 mVrms/Vs)
Table 13. Measurement time mode = MED, LONG
Test signal voltage
Test
5 mVrms -
50 mVrms -
0.3 Vrms -
1 Vrms -
10 Vrms -
frequency [Hz] 50 mVrms
0.3 Vrms
1 Vrms
10 Vrms
20 Vrms
20 - 100
(0.25%) ×
0.25%
0.10%
0.15%
0.15%
(30
mVrms/Vs)
100 - 1 M
(0.1%) ×
0.10%
0.05%
0.10%
0.15%
(30
mVrms/Vs)
1 M - 2 M
(0.2%) ×
0.20%
0.10%
0.20%
0.30%
(30
mVrms/Vs)
Vs [Vrms]