Measurement time

Definition

This is the time between the trigger and the end of measurement (EOM) output on the
handler interface.

Conditions

Table 53 shows the measurement time when the following conditions are satisfied:

• Normal impedance measurement other than Ls-Rdc, Lp-Rdc, Vdc-Idc
• Impedance range mode: hold range mode
• DC bias voltage level monitor: OFF
• DC bias current level monitor: OFF
• Trigger delay: 0 s
• Step delay: 0 s
• Calibration data: OFF
• Display mode: blank

Table 53. Measurement time [ms](DC bias:OFF)

Measurement

time mode

Test frequency

20 Hz

100 Hz

1 kHz

10 kHz

100 kHz

1 MHz

2 MHz

1 LONG

480 300

240 230

220

220

220

2 MED

380 180

110 92

89

88

88

3 SHORT

330 100

20

7.7

5.7

5.6

5.6

Figure 9. Measurement time (DC bias: OFF)

20

100

1k

10k

100k

1M 2M

0.01

0.001

0.1

1

10

Test frequency [Hz]

Measurement time [sec]

1. LONG

2. MED

3. SHORT