beautypg.com

Atec Agilent-E4980A User Manual

Page 12

background image

12

Effect by impedance of DUT

Table 14. For impedance of DUT below 30 Ω, the following value is added.

Test

Impedance of DUT

frequency [Hz]

1.08 Ω ≤ |Zx| < 30 Ω

|Zx| < 1.08 Ω

20 - 1 M

0.05%

0.10%

1 M - 2 M

0.10%

0.20%

Table 15. For impedance of DUT over 9.2 k Ω, the following value is added.

Test

Impedance of DUT

frequency [Hz]

9.2 kΩ < |Zx| ≤ 92 kΩ

92 kΩ < |Zx|

10 k - 100 k

0%

0.05%

100 k - 1 M

0.05%

0.05%

1 M - 2 M

0.10%

0.10%

Effect of cable extension

When the cable is extended, the following element is added per one meter.

0.015 % × (Fm/1 MHz)

2

× (L_cable)

2

Fm [Hz]

Test Frequency

L_cable [m]

Cable length