Settling time

Table 50. Test frequency setting time

Test frequency setting time

Test frequency (Fm)

5 ms

Fm ≥ 1 kHz

12 ms

1 kHz > Fm ≥ 250 Hz

22 ms

250 Hz > Fm ≥ 60 Hz

42 ms

60 Hz > Fm

Table 51. Test signal voltage setting time

Test signal voltage setting time

Test frequency (Fm)

11 ms

Fm ≥ 1 kHz

18 ms

1 kHz > Fm ≥ 250 Hz

26 ms

250 Hz > Fm ≥ 60 Hz

48 ms

60 Hz > Fm

Switching of the impedance range is as follows:

≤ 5 ms/ range switching

Measurement circuit protection

The maximum discharge withstand voltage, where the internal circuit remains protected
if a charged capacitor is connected to the UNKNOWN terminal, is given below.

Table 52. Maximum discharge withstand voltage

Maximum discharge withstand voltage Range of capacitance value C of DUT

1000 V

C < 2 µF

√ 2/C V

2 µF ≤ C

Figure 8. Maximum discharge withstand voltage

Supplemental
Information

Note

Discharge capacitors before connecting them
to the UNKNOWN terminal or a test fixture to
avoid damages to the instrument.