Test signal – Atec Agilent-E4980A User Manual

Page 4

background image

Table 1. Trigger delay time

Range

0 s - 999 s

Resolution

100 µs (0 s - 100 s)

1 ms (100 s - 999 s)

Table 2. Step delay time

Range

0 s - 999 s

Resolution

100 µs (0 s - 100 s)

1 ms (100 s - 999 s)

Measurement terminal: Four-terminal pair

Test cable length: 0 m, 1 m, 2 m, 4 m

Measurement time modes: Short mode, medium mode, long mode.

Table 3. Averaging

Range

1 - 256 measurements

Resolution

1

Test signal

Table 4. Test frequencies

Test frequencies

20 Hz - 2 MHz

Resolution

0.01 Hz (20 Hz - 99.99 Hz)

0.1 Hz (100 Hz - 999.9 Hz)

1 Hz (1 kHz - 9.999 kHz)

10 Hz (10 kHz - 99.99 kHz)

100 Hz (100 kHz - 999.9 kHz)

1 kHz (1 MHz - 2 MHz)

Measurement accuracy

± 0.01%

Table 5. Test signal modes

Normal

Program selected voltage or current at the measurement

terminals when they are opened or short-circuited, respectively.

Constant

Maintains selected voltage or current at the device under test

(DUT) independently of changes in impedance of DUT.

4