Effect of cable length

Table 21. When the cable is extended, multiply Yo by the following factor.

Test

Cable length

frequency [Hz]

0 m

1 m

2 m

4 m

100 - 100 k

1

1 + 5 × Fm/1 MHz

1 + 10 × Fm/1 MHz

1 + 20 × Fm/1 MHz

100 k - 1 M

1

1 + 0.5 × Fm/1 MHz

1 + 1 × Fm/1 MHz

1 + 2 × Fm/1 MHz

1 M - 2 M

1

1 + 1 × Fm/1 MHz

1 + 2 × Fm/1 MHz

1 + 4 × Fm/1 MHz

Fm [Hz]

Test frequency

Temperature factor Kt

Table 22. The temperature factor Kt is given below.

Temperature [

°

C]

Kt

0 - 18

4

18 - 28

1

28 - 55

4