Definitions how to use tables – Atec Agilent-E4980A User Manual
Page 2
2
All specifications apply to the conditions of a 0 to 55 °C temperature range, unless
otherwise stated, and 30 minutes after the instrument has been turned on.
Specifications (spec.): Warranted performance. Specifications include guardbands to
account for the expected statistical performance distribution, measurement uncertainties,
and changes in performance due to environmental conditions.
Supplemental information is provided as information that is useful in operating the
instrument, but is not covered by the product warranty. This information is classified as
either typical or nominal.
Typical (typ.): Expected performance of an average unit without taking guardbands
into account.
Nominal (nom.): A general descriptive term that does not imply a level of performance.
When measurement conditions fall under multiple categories in a table, apply the best
value.
For example, basic accuracy Ab is 0.10% under the following conditions;
Measurement time mode
SHORT
Test frequency
125 Hz
Test signal voltage
0.3 Vrms
Definitions
How to Use Tables