DC resistance (Rdc) accuracy

Absolute measurement accuracy Aa

Absolute measurement accuracy Aa is given as

Equation 15.

Aa = Ae + Acal

Aa

Absolute accuracy (% of reading value)

Ae

Relative accuracy (% of reading value)

Acal Calibration accuracy

Relative measurement accuracy Ae

Relative measurement accuracy Ae is given as

Equation 16.

Ae = [Ab + (Rs /|Rm|+ Go × |Rm|) Ч 100 ] Ч Kt

Rm Measurement

value

Ab Basic

accuracy

Rs

Short offset [Ω]

Go Open

offset

[S]

Kt

Temperature coefficient

Calibration accuracy Acal

Calibration accuracy Acal is 0.03%.

Basic accuracy Ab

Table 42. Basic accuracy Ab is given below.

Measurement

Test signal voltage

time mode

≤ 2 Vrms

> 2 Vrms

SHORT

1.00%

2.00%

MED

0.30%

0.60%

Open offset Go

Table 43. Open offset Go is given below.

Measurement

Test signal voltage

time mode

≤ 2 Vrms

> 2 Vrms

SHORT

50 nS

500 nS

MED

10 nS

100 nS

Short offset Rs

Table 44. Short offset Rs is given below.

Measurement

Test signal voltage

time mode

≤ 2 Vrms

> 2 Vrms

SHORT

25 mΩ

250 mΩ

MED

5 mΩ

50 mΩ