Atec Agilent-E4980A User Manual
Page 22
22
Table 36. Test signal voltage ≤ 0.2 Vrms (measurement time mode = MED, LONG)
DC bias
Impedance range [Ω]
current range
< 100
100
300, 1 k
3 k, 10 k
30k, 100 k
20 µA
150 µA
30 µA
3 µA
300 nA
45 nA
200 µA
150 µA
30 µA
3 µA
300 nA
300 nA
2 mA
150 µA
30 µA
3 µA
3 µA
3 µA
20 mA
150 µA
30 µA
30 µA
30 µA
30 µA
100 mA
150 µA
150 µA
150 µA
150 µA
150 µA
Table 37. 0.2 Vrms < test signal voltage ≤ 2 Vrms
(measurement time mode = MED, LONG)
DC bias
Impedance range [Ω]
current range
< 100
100, 300
1k, 3 k
10k, 30 k
100 k
20 µA
150 µA
30 µA
3 µA
300 nA
45 nA
200 µA
150 µA
30 µA
3 µA
300 nA
300 nA
2 mA
150 µA
30 µA
3 µA
3 µA
3 µA
20 mA
150 µA
30 µA
30 µA
30 µA
30 µA
100 mA
150 µA
150 µA
150 µA
150 µA
150 µA
Table 38. Test signal voltage > 2 Vrms (measurement time mode = MED, LONG)
DC bias
Impedance range [Ω]
current range
≤ 300
1 k, 3 k
10k, 30 k
100 k
20 µA
150 µA
30 µA
3 µA
300 nA
200 µA
150 µA
30 µA
3 µA
300 nA
2 mA
150 µA
30 µA
3 µA
3 µA
20 mA
150 µA
30 µA
30 µA
30 µA
100 mA
150 µA
150 µA
150 µA
150 µA
Table 39. Input impedance (nominal)
Input
impedance Conditions
0 Ω
Other than conditions below.
20 Ω
Test signal voltage ≤ 0.2 Vrms, Impedance range ≥ 3 k Ω,
DC bias current range ≤ 200 µA
Test signal voltage ≤ 2 Vrms, Impedance range ≥ 10 kΩ,
DC bias current range ≤ 200 µA
Test signal voltage > 2 Vrms, Impedance range = 100 kΩ,
DC bias current range ≤ 200 µA
DC source signal
Table 40. Test signal voltage
Range
–10 V to 10 V
Resolution
1 mV
Accuracy
0.1% + 3 mV (23
°
C ± 5
°
C)
(0.1% + 3 mV) x 4
(0
to
18
°
C or 28 to 55
°
C)
Table 41. Test signal current
Range
–45 mA to 45 mA (nominal)
Output impedance
100 Ω (nominal)