Table 36. Test signal voltage ≤ 0.2 Vrms (measurement time mode = MED, LONG)

DC bias

Impedance range [Ω]

current range

< 100

100

300, 1 k

3 k, 10 k

30k, 100 k

20 µA

150 µA

30 µA

3 µA

300 nA

45 nA

200 µA

150 µA

30 µA

3 µA

300 nA

300 nA

2 mA

150 µA

30 µA

3 µA

3 µA

3 µA

20 mA

150 µA

30 µA

30 µA

30 µA

30 µA

100 mA

150 µA

150 µA

150 µA

150 µA

150 µA

Table 37. 0.2 Vrms < test signal voltage ≤ 2 Vrms
(measurement time mode = MED, LONG)

DC bias

Impedance range [Ω]

current range

< 100

100, 300

1k, 3 k

10k, 30 k

100 k

20 µA

150 µA

30 µA

3 µA

300 nA

45 nA

200 µA

150 µA

30 µA

3 µA

300 nA

300 nA

2 mA

150 µA

30 µA

3 µA

3 µA

3 µA

20 mA

150 µA

30 µA

30 µA

30 µA

30 µA

100 mA

150 µA

150 µA

150 µA

150 µA

150 µA

Table 38. Test signal voltage > 2 Vrms (measurement time mode = MED, LONG)

DC bias

Impedance range [Ω]

current range

≤ 300

1 k, 3 k

10k, 30 k

100 k

20 µA

150 µA

30 µA

3 µA

300 nA

200 µA

150 µA

30 µA

3 µA

300 nA

2 mA

150 µA

30 µA

3 µA

3 µA

20 mA

150 µA

30 µA

30 µA

30 µA

100 mA

150 µA

150 µA

150 µA

150 µA

Table 39. Input impedance (nominal)

Input
impedance Conditions

0 Ω

Other than conditions below.

20 Ω

Test signal voltage ≤ 0.2 Vrms, Impedance range ≥ 3 k Ω,

DC bias current range ≤ 200 µA

Test signal voltage ≤ 2 Vrms, Impedance range ≥ 10 kΩ,

DC bias current range ≤ 200 µA

Test signal voltage > 2 Vrms, Impedance range = 100 kΩ,

DC bias current range ≤ 200 µA

DC source signal

Table 40. Test signal voltage

Range

–10 V to 10 V

Resolution

1 mV

Accuracy

0.1% + 3 mV (23

°

C ± 5

°

C)

(0.1% + 3 mV) x 4

(0

to

18

°

C or 28 to 55

°

C)

Table 41. Test signal current

Range

–45 mA to 45 mA (nominal)

Output impedance

100 Ω (nominal)