Table 54. Measurement time when option 005 is installed [ms]
(DC bias: OFF)

Measurement

time mode

Test frequency

20 Hz

100 Hz

1 kHz

10 kHz

100 kHz

1 MHz

2 MHz

1 LONG

1190 650

590 580

570

570

570

2 MED

1150 380

200 180

180

180

180

3 SHORT

1040 240

37

25

23

23

23

Figure 10. Measurement time (DC bias: OFF, Option 005)

When DC bias is ON, the following time is added:

Table 55. Additional time when DC bias is ON [ms]

Test frequency
20 Hz

100 Hz

1 kHz

10 kHz

100 kHz

1 MHz

2 MHz

30 30 10 13 2

0.5 0.5

When the number of averaging increases, the measurement time is given as

Equation 17.

MeasTime + (Ave – 1) × AveTime

MeasTime

Measurement time calculated based on Table 53 and Table 54

Ave

Number of averaging

AveTime

Refer to Table 56

Table 56. Additional time per averaging [ms]

Measurement
time mode

Test frequency

20 Hz

100 Hz

1 kHz

10 kHz

100 kHz

1 MHz

2 MHz

SHORT 51 11 2.4

2.3 2.3 2.2 2.2

MED 110

81

88

87

85

84 84

LONG 210

210

220

220

220

210

210

20

100

1k

10k

100k

1M 2M

0.01

0.001

0.1

1

10

Test frequency [Hz]

M

easuremen

t

ti

me

[sec

]

1. LONG

2. MED

3. SHORT