Atec Agilent-E4980A User Manual

Page 5

background image

5

Signal level

Table 6. Test signal voltage

Range

0 Vrms - 2.0 Vrms

Resolution

100 µVrms (0 Vrms - 0.2 Vrms)

200 µVrms (0.2 Vrms - 0.5 Vrms)

500 µVrms (0.5 Vrms - 1 Vrms)

1 mVrms (1 Vrms - 2 Vrms)

Accuracy

Normal

±(10% + 1 mVrms) Test frequency ≤ 1 MHz: spec.

Test frequency > 1 MHz: typ.

Constant

1

±(6% + 1 mVrms) Test frequency ≤ 1 MHz: spec.

Test frequency > 1 MHz: typ.

Table 7. Test signal current

Range

0 Arms - 20 mArms

Resolution

1 µArms (0 Arms - 2 mArms)

2 µArms (2 mArms - 5 mArms)

5 µArms (5 mArms - 10 mArms)

10 µArms (10 mArms - 20 mArms)

Accuracy

Normal

±(10% + 10 µArms) Test frequency ≤ 1 MHz: spec.

Test frequency > 1 MHz: typ.

Constant

1

±(6% + 10 µArms) Test frequency < = 1 MHz: spec.

Test frequency > 1 MHz: typ.

Output impedance: 100 Ω (nominal)

Test signal level monitor function

• Test signal voltage and test signal current can be monitored.
• Level monitor accuracy:

Table 8. Test signal voltage monitor accuracy (Vac)

Test signal voltage

2

Test

frequency Specification

5 mVrms - 2 Vrms

≤ 1 MHz

± (3% of reading value + 0.5 mVrms)

> 1 MHz

± (6% of reading value + 1 mVrms)

Table 9. Test signal current monitor accuracy (lac)

Test signal current

2

Test

frequency Specification

50 µArms - 20 mArms ≤ 1 MHz

± (3% of reading value + 5 µArms)

> 1 MHz

± (6% of reading value + 10 µArms)

1. When auto level control function is on.
2. This is not an output value but rather a displayed

test signal level.